Comparison of Four Dual-Energy CT Scanner Technologies for Determining Renal Stone Composition: A Phantom Approach
Ali Pourvaziri1, Anushri Parakh1, Jinjin Cao1
1From the Department of Radiology (A. Pourvaziri, J.C., A.K.), Harvard Catalyst Biostatistics Consulting Unit (J.L.), and Department of Urology (D.S.), Massachusetts General Hospital, 55 Fruit St, Boston, MA 02114; Department of Radiology, Brigham and Women's Hospital, Boston, Mass (A. Parakh); and Department of Radiology, University of Washington, Seattle, Wash (B.E.).
Dual-energy CT (DECT) technologies accurately determine renal stone composition using standard and low doses. Performance varied by scanner, technique, and stone type, but all four tested DECT systems showed effectiveness.
Area of Science:
- Radiology
- Medical Imaging
- Urology
Background:
- Dual-energy CT (DECT) is valuable for assessing renal stone composition.
- Limited comparative data exist for multivendor DECT technologies.
- This study addresses the need for performance comparison across different DECT systems.
Purpose of the Study:
- To compare the diagnostic performance of four DECT technologies.
- To evaluate DECT for renal stone composition analysis at standard and low radiation doses.
- To assess the impact of different acquisition protocols on DECT performance.
Main Methods:
- In vitro phantom study with 71 urinary stones of known composition.
- Scans performed on four DECT technologies: rapid kilovolt peak switching DECT (rsDECT), dual-source DECT, twin-beam DECT (tbDECT), and dual-layer detector-based CT (dlDECT).
- Analysis using effective atomic number and dual-energy ratio indexes; performance assessed via area under the receiver operating characteristic curve (AUC).
Main Results:
- All DECT technologies, except tbDECT, achieved AUC > 0.95 for differentiating urate stones.
- Moderate accuracy (AUC: 0.70-0.83) was observed for differentiating calcium oxalate monohydrate stones.
- Brushite stones were differentiated from urate stones with AUC > 0.99.
- No significant performance difference was found between standard- and low-dose acquisitions or between vendor-recommended and dose-optimized protocols.
Conclusions:
- All four DECT technologies effectively determine renal stone composition at both standard and low doses.
- Performance is influenced by scanner parameters, DECT technique, and specific stone composition.
- DECT is a reliable tool for non-invasive renal stone characterization.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Imaging Studies for Cardiovascular System V: CT
Urinary Tract Calculi III: Medical Management
Imaging Studies IV: Magnetic Resonance Imaging