1Departamento de Física, Faculdade de Filosofia, Ciências e Letras de Ribeirão Preto, Universidade de São Paulo, Ribeirão Preto, 14040-900, São Paulo, Brazil.
Abstract:
Metrobras, a private company in Brazil, is developing a new laboratory for the calibration of neutron detectors in collaboration with the National Laboratory for Metrology of Ionizing Radiation (LNMRI). This work aimed at assessing neutron spectra, ambient dose equivalent rates, and neutron scattering in the calibration room of Metrobras Neutron Detectors Calibration Laboratory (LCDNM) using Monte Carlo simulation with MCNP5. Three models were simulated: vacuum, air, and the complete with the several materials employed in the actual room. Neutron spectra as a function of the distance from the source were obtained, compared, and used to calculate ambient dose equivalent at different points in the calibration room and results were compared to other neutron laboratories. Results show that LCDNM ambient dose equivalent rates are comparable to those of other neutron laboratories with 531.2 μSv . h-1 at 100 cm from the 592 GBq activity 241AmBe source. This work shows that LCDNM may be used as a calibration laboratory for neutron detectors in accordance with ISO 8529.
More Related Videos
14:11Quantification of Hydrogen Concentrations in Surface and Interface Layers and Bulk Materials through Depth Profiling with Nuclear Reaction Analysis
Published on: March 29, 2016
11:25In Situ Neutron Powder Diffraction Using Custom-made Lithium-ion Batteries
Published on: November 10, 2014
Related Concept Videos
Nuclear Transmutation
Calorimetry
Constant Pressure Calorimetry
Maxwell-Boltzmann Distribution: Problem Solving
This distribution function f(v) is defined by saying that the expected number N (v1,v2) of particles with speeds between v1 and v2 is given by
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Nuclear Fission