Three-dimensional chemical exchange saturation transfer imaging using compressed SENSE for full z-spectrum
Tatsuhiro Wada1, Chiaki Tokunaga2, Osamu Togao3
1Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Japan; Department of Health Sciences, Graduate school of medical Sciences, Kyushu University, Japan.
Purpose:
To evaluate the accuracy of three-dimensional (3D) chemical exchange saturation transfer (CEST) imaging with a compressed sensing (CS) and sensitivity encoding (SENSE) technique (CS-SENSE) for full z-spectrum acquisition.
Methods:
All images were acquired on 3-T magnetic resonance imaging (MRI) scanner. In the phantom study, we used the acidoCEST imaging. The phantoms were prepared in seven vials containing different concentrations of iopamidol mixed in phosphate-buffered solution with different pH values. The CEST ratios were calculated from the two CEST effects. We compared the CEST ratios obtained with three different 3D CEST imaging protocols (CS-SENSE factor 5, 7, 9) with those obtained with the 2D CEST imaging as a reference standard. In the clinical study, 21 intracranial tumor patients (mean 49.7 ± 17.2 years, 7 males and 14 females) were scanned. We compared the intratumor magnetization transfer ratio asymmetry (MTRasym) obtained with 3D CEST imaging with those obtained with 2D CEST imaging as a reference standard.
Results:
A smaller CS-SENSE factor resulted in higher agreement and better correlations with the 2D CEST imaging in the phantom study (CS-SENSE 5; ICC = 0.977, R2 = 0.8943, P < 0.0001: CS-SENSE 7; ICC = 0.970, R2 = 0.9013, P < 0.0001: CS-SENSE 9; ICC = 0.934, R2 = 0.8156 P < 0.0001). In the brain tumors, the means and percentile values of MTRasym at 2.0 and 3.5 ppm showed high linear correlations (R2 = 0.7325-0.8328, P < 0.0001) and high ICCs (0.859-0.907), which enabled successful multi-slice CEST imaging.
Conclusions:
The 3D CEST imaging with CS-SENSE provided equivalent contrast to 2D CEST imaging; moreover, a z-spectrum with a wide scan range could be obtained.
Related Concept Videos
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Two-Dimensional (2D) NMR: Overview
The first step is the preparation period, during which nucleus A is excited with a radiofrequency pulse....
2D NMR: Overview of Homonuclear Correlation Techniques
COSY90 is the standard two-dimensional (2D) COSY experiment that...
2D NMR: Heteronuclear Single-Quantum Correlation Spectroscopy (HSQC)
Tandem Mass Spectrometry
Secondary fragmentations occur in the interaction cell and can be induced by various factors. Fragmentation induced by collision with inert gases, such as N2, Ar, He, etc., is called collision-induced...