Related Experiment Video
Updated: Sep 21, 2025

A Miniaturized Glycan Microarray Assay for Assessing Avidity and Specificity of Influenza A Virus Hemagglutinins
Published on: May 29, 2016
Quick and ultra-sensitive digital assay of influenza virus using sub-picoliter microwells
Hiroki Ashiba1, Masato Yasuura1, Takashi Fukuda1
1Sensing System Research Center, National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba, Japan.
Abstract:
Quick and sensitive virus detection methods have been developed using sub-picoliter microwells. One method uses an aggregation-induced emission (AIE) reagent, and the other uses an enzymatic reaction supported with immunomagnetic beads at a high concentration of 108 particles/mL. Examination of influenza A virus detection using the AIE reagent exhibited a detection limit of 3 × 105 copies/mL, which was achievable within 1 min of the total measurement time. The high-concentration immunomagnetic beads method exhibited a detection limit as low as 1 × 102 copies/mL. The developed methods are effective and practical tools for ultrafast and ultrasensitive virus detection.