Related Experiment Video
Updated: Sep 21, 2025

Attaching Biological Probes to Silica Optical Biosensors Using Silane Coupling Agents
Published on: May 1, 2012
Simple preparation of surface molecularly imprinted polymer based on silica particles for trace level assay of
S Irem Kaya1,2, M Emin Corman1,3, Lokman Uzun4
1Faculty of Pharmacy, Department of Analytical Chemistry, Ankara University, Ankara, Turkey.
Abstract:
A new electrochemical sensor based on molecularly imprinted tetraethyl orthosilicate (TEOS)-based porous interface was developed for selective recognition of bisphenol F (BPF) in this study. The sensor was prepared by depositing the solution containing TEOS and L-tryptophan (L-Trp) in the presence of cetyltrimethylammonium bromide (CTAB) as a pore-maker via hydrolysis/condensation reaction on the glassy carbon electrode (GCE). While the surface morphology and structure characterization were carried out using Fourier transform infrared spectroscopy (FTIR) and scanning electron microscopy (SEM), electrochemical characterization was performed through electrochemical impedance spectroscopy (EIS) and cyclic voltammetry (CV). The resulted MIP(TEOS:L-Trp)@GCE achieved a wide linear range of 1 × 10-15-1 × 10-14 M for BPF detection with an excellent detection limit of 0.291 fM. Furthermore, the recovery of BPF from spiked bottled water and serum samples varied between 98.83 and 101.03%. These results demonstrate that MIP(TEOS:L-Trp)@GCE was found to be a simple, sensitive, and selective smart interface to detect trace pollution even from complicated samples.