Related Experiment Video
Updated: Sep 21, 2025

Erythrocyte Sedimentation Rate: A Physics-Driven Characterization in a Medical Context
Published on: March 24, 2023
Effect of storage temperature and time on erythrocyte sedimentation rate
Qi-Lei Hu1,2, Zuo-Jie Li3, Li Lin4
1Department of Clinical Laboratory, Linping Campus, The Second Affiliated Hospital of Zhejiang University School of Medicine, Hangzhou, 311100, Zhejiang, China.
Objective:
This paper explores the effect of blood sample storage temperature and time on the erythrocyte sedimentation rate (ESR) by using the Weiss method.
Methods:
Whole blood samples were collected from 80 patients and diluted 1:9 with sodium citrate solution. Each sample was split into two tubes. Using the Weiss method, ESR was tested within 1 h of collection, and one sample was placed at 4 °C and the other at room temperature (23 ± 2 °C). ESR was then measured at 2, 4, 6, 8, 12, and 24 h. The data were statistically analyzed with consideration for temperature and time.
Results:
ESR decreased gradually over 6 h at room temperature, but the results were not statistically significant. Similarly, there was no significant difference in the decline of ESR within 8 h at 4 °C. However, ESR results decreased significantly after the samples were stored at room temperature for more than 6 h or at 4 °C for more than 8 h. ESR reduction was lower in the samples stored at 4 °C than in those stored at room temperature over the same time period.
Conclusion:
Blood sample storage temperature and duration can affect the measurement of ESR using the Weiss method. ESR testing should be completed within 4 h of sample collection in clinical work.
Related Concept Videos
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature...
Effect of Temperature Change on Reaction Rate
Sample Handling
Samples should be transported carefully from collection points to the laboratory. They should be properly sealed and clearly labeled to prevent cross-contamination. To preserve the sample integrity, optimal temperature conditions during transport are essential. This could involve using...