A method for continuous rhythm classification and early detection of ventricular fibrillation during CPR
Heemun Kwok1, Jason Coult2, Jennifer Blackwood3
1Department of Emergency Medicine, University of Washington, Seattle, WA, United States; Center for Progress in Resuscitation, University of Washington, Seattle, WA, United States.
Aim:
We developed a method which continuously classifies the ECG rhythm during CPR in order to guide clinical care.
Methods:
We conducted a retrospective study of 432 patients treated following out-of-hospital cardiac arrest. Continuous ECG sequences from two-minute CPR cycles were extracted from defibrillator recordings and further divided into five-second clips. We developed an algorithm using wavelet analysis, hidden semi-Markov modeling, and random forest classification. The algorithm classifies individual clips as asystole, organized rhythm, ventricular fibrillation, or Inconclusive while integrating information from previous clips. Classifications were compared to manual annotations to estimate accuracy in an independent validation dataset. Continuous sequences were classified as shockable, non-shockable, or Inconclusive; classifications were used to compute shock sensitivity and specificity.
Results:
Of 432 patient-cases, 290 were used for development and 142 for validation. In the 12,294 validation ECG clips during CPR, accuracies were 0.88 (95% CI 0.85-0.91) for asystole, 0.98 (95% CI 0.98-0.99) for organized rhythm, and 0.97 (95% CI 0.96-0.97) for ventricular fibrillation, with 43% classified as Inconclusive. Of 457 continuous sequences, shock sensitivity was 0.90 (95% CI 0.86-0.93), shock specificity was 0.98 (95% CI 0.93-0.99), and 7% were Inconclusive. Median delay to ventricular fibrillation recognition was 10 (IQR 5-32) seconds.
Conclusion:
A novel algorithm continuously classified the primary resuscitation rhythms-asystole, organized rhythms, and ventricular fibrillation-with 88-98% accuracy, enabling accurate shock advisory guidance during most two-minute CPR cycles. Additional investigation is required to understand how algorithm implementation could affect rescuer actions and clinical outcomes.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Cardiopulmonary Resuscitation III: AED Use
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias III: Characteristics of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Cardiopulmonary Resuscitation IV: Pharmacological Management