[A Study of Longitudinal NPS Measurement in CT Images Based on the Central Cross-section Theorem]
Akihiro Narita1, Masaki Ohkubo1, Yuki Ohsugi1
1Department of Radiological Technology, Graduate School of Health Sciences, Niigata University.
Purpose:
Various approaches in noise power spectrum (NPS) analysis are currently used for measuring a patient's longitudinal (z-direction) NPS from three-dimensional (3D) CT volume data. The purpose of this study was to clarify the relationship between those NPSs and 3D-NPS based on the central slice theorem.
Methods:
We defined the 3D-NPS(fx, fy, fz) that was calculated by 3D Fourier transform (FT) from 3D noise data (3D-Noise(x, y, z), x-y scan plane). Here, fx, fy and fz are spatial frequencies corresponding to the axes of x, y and z, respectively. Based on the central slice theorem, we described three relationships as follows. (1) The fz-directional NPS calculated from the 3D-Noise(x=0, y=0, z) is equal to the profile obtained by projecting 3D-NPS(fx, fy, fz) in fx- and fy-directions. (2) The fz-directional NPS calculated from the profile obtained by projecting 3D-Noise(x=0, y, z) in the y-direction is equal to the profile at fy=0 in the data obtained by projecting 3D-NPS(fx, fy, fz) in the fx-direction. (3) The fz-directional NPS calculated from the profile obtained by projecting 3D-Noise(x, y, z) in x and y-directions is equal to the profile of 3D-NPS(fx=0, fy=0, fz). To verify them, we compared the NPSs measured from actual 3D noise data that were obtained using a cylindrical water phantom.
Results:
In each relationship (1)-(3), the fz-directional NPS matched the profile obtained from the 3D-NPS(fx, fy, fz).
Conclusion:
Based on the central slice theorem, we clarified the relationships between fz-directional NPSs and 3D-NPS. We should understand them and then consider which method should be used for fz-directional NPS measurement.
More Related Videos
13:07Optical Coherence Tomography Based Biomechanical Fluid-Structure Interaction Analysis of Coronary Atherosclerosis Progression
Published on: January 15, 2022
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Deformations in a Transverse Cross Section
As the material stretches, it expands or contracts in orthogonal directions to the load. This phenomenon varies...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Profile Leveling and Cross Sections