A disposable printed amperometric biosensor for clinical evaluation of creatinine in renal function detection
Yaping Dong1, Xiaojin Luo2, Yiqun Liu2
1Department of Medicine, Renal Division, Hospital 1, Peking University, Beijing, 100034, PR China; School of Materials Science and Engineering, Peking University, Beijing, 100871, PR China.
Abstract:
In clinical practice, sera creatinine level is regarded as a crucial biomarker for the diagnosis, staging and monitoring of kidney disease. An amperometric biosensor is rapid, accurate, and cost-effective, with a portability and a simple operation. Herein, we report for the firsttime a disposable, printed amperometric biosensor for the clinical evaluation of creatinine in renal function detection. The sensor is constructed based on Prussian blue/carbon-graphite paste as the working electrode and the immobilization of creatinine amidohydrolase, creatine amidinohydrolase and sarcosine oxidase. The creatinine biosensor shows a linear detection range from 0.05 to 1.4 mM with a detection time of about 3 min. In addition, the sensor shows a high stability that can maintain above 86% of the initial activity after being stored for over 4 months. Moreover, the sensor shows almost the same results as those with the Jaffe method for measuring the real blood samples. We anticipate that the creatinine biosensor could be widely used in the medical and healthcare areas, especially for at-home testing and onsite medical examinations.