Related Experiment Video
Updated: Sep 20, 2025

Intraoperative Assessment of Resection Margins in Oral Cavity Cancer: This is the Way
Published on: May 10, 2021
Ex vivo confocal microscopy for surgical margin assessment: A histology-compared study on 109 specimens
1Department of Dermatology and Venereology, Hôpital de Beaumont Lausanne University Hospital Center Lausanne Switzerland.
Background:
The assessment of surgical margins is mandatory to prevent local recurrence or distant dissemination of skin cancers. Histological examination of haematoxylin and eosin (H&E)-stained slides from paraffin-embedded or frozen samples is the gold standard for margin assessment, but is a time-consuming procedure. Ex vivo confocal laser scanning microscopy (CLSM) is an upcoming technique that scans unfixed fresh tissue rapidly, allowing fast per-operative margin assessment.
Objective:
Here, we propose to assess the efficiency of a new ex vivo confocal microscope for the per-operative assessment of surgical margins.
Methods:
We analyzed 16 biopsies and 93 surgical specimens of basal cell and squamous cell carcinomas by ex vivo CLSM using Histolog® Scanner V2. Surgical specimens included fusiform excisions, slow-Mohs peripheral and deep compartments, and Mohs excisions. The time required from surgical excision to image analysis was recorded and the quality of the images obtained for each specimen assessed. The presence or absence of tumour was estimated based on ex vivo CLSM images and compared with conventional H&E-stained sections from paraffin-embedded or frozen (Mohs) specimens.
Results:
Mean time for specimen processing using Histolog Scanner was 5.1 ± 3.4 min. We obtained 89% of high quality images. Mean time for confocal image analysis was 1 ± 0.76 min. The diagnostic sensitivity and specificity for ex vivo CLSM compared to classical H&E procedures were respectively 93% and 100% when performed on tumour biopsies. The overall sensitivity and specificity for ex vivo CLSM for margin assessment compared to classical H&E procedures were respectively 61.5% and 95%, with variations depending on the type of tumour or surgical specimen analyzed. In particular, we obtained 80% sensitivity and 100% specificity for the assessment of BCC surgical margins.
Conclusion:
Our data suggest that ex vivo CLSM using Histolog® Scanner V2 could be a valid help for surgeons for a fast and accurate per-operative margin analysis.
More Related Videos
07:47Author Spotlight: 3D Scanning and Augmented Reality for Enhanced Cancer Surgery Communication
Published on: December 15, 2023
14:21Optical Frequency Domain Imaging of Ex vivo Pulmonary Resection Specimens: Obtaining One to One Image to Histopathology Correlation
Published on: January 22, 2013