Evaluation of automatic tube current modulation of CT scanners using a dedicated and the CTDI dosimetry phantoms
Ioannis A Tsalafoutas1, Shady AlKhazzam1, Huda AlNaemi1,2
1Medical Physics Section, OHS Department, Hamad Medical Corporation, Doha, Qatar.
Purpose:
To investigate the operation principles of the automatic tube current modulation (ATCM) of a CT scanner, using a dedicated phantom and the CT dosimetry index (CTDI) phantom.
Material And Methods:
The Mercury 4.0 phantom and three different configurations of the CTDI dosimetry phantom were employed. A frequently used clinical scanning protocol was employed as a basis for the acquisitions performed with all phantoms, using both scanning directions. Additional acquisitions with different pitch and examination protocols were performed with Mercury phantom, to further explore their effect on ATCM and the resulting image quality. Different software named DICOM Info Extractor, ImageJ, and imQuest, were used to derive CTDIvol and table position, image noise, and water equivalent diameter (WED) of each phantom CT image, respectively. ImQuest was also used to derive the detectability index (d') of five different materials (air, solid water, polystyrene, iodine, and bone) embedded in the Mercury phantom.
Results:
It was exhibited with all four phantoms that the scanning direction greatly affects the modulation curves. The fitting of the dose modulations curves suggested that for each table position what determines the CTDIvol value is the WED values of the phantom structures laying ahead towards the scanning direction, for a length equal to the effective width of the X-ray beam. Furthermore, it was also exhibited that ATCM does not fully compensate for larger thicknesses, since images of larger WED phantom sections present more noise (larger SD) in all four phantoms and in Mercury 4.0 phantom smaller detectability (d').
Conclusion:
Mercury 4.0 is a dedicated phantom for a complete and in-depth evaluation of the ATCM operation and the resulting image quality. However, in its absence, different CTDI configurations can be used as an alternative to investigate and comprehend some basic operation principles of the CT scanners' ATCM systems.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT