Analysis for carboxyhemoglobin by gas chromatography and multicomponent spectrophotometry compared
Clinical Chemistry
|May 1, 1987
Summary
Gas chromatography (GC) accurately measures carboxyhemoglobin (HbCO) compared to multicomponent spectrophotometric analysis (MCA), especially for higher HbCO levels. This method is reliable for HbCO analysis and blood sample stability.
Area of Science:
- Clinical Chemistry
- Analytical Chemistry
Background:
- Carboxyhemoglobin (HbCO) is a critical biomarker in clinical diagnostics.
- Accurate measurement of HbCO is essential for patient diagnosis and monitoring.
Purpose of the Study:
- To compare the accuracy of gas chromatography (GC) for measuring carboxyhemoglobin (HbCO) against multicomponent spectrophotometric analysis (MCA).
- To assess the linearity and stability of the GC method for HbCO determination.
Main Methods:
- Comparison of HbCO measurements obtained by GC and MCA across various concentration ranges.
- Evaluation of method linearity from 0.15% to 30% HbCO.
- Assessment of sample stability after storage at 4°C for two months.
Main Results:
- Good correlation between GC and MCA was observed for HbCO ranges above 1.5% (r = 0.93–0.99).
- Lower correlation was noted in the clinically normal range of 0–1.5% (r = 0.87).
- The GC method demonstrated linearity from 0.15% to 30% HbCO, and samples were stable for two months at 4°C.
Conclusions:
- Gas chromatography provides a reliable and accurate method for quantifying carboxyhemoglobin, particularly at higher concentrations.
- The GC method is suitable for clinical use due to its linearity and the stability of blood samples under refrigerated storage.