Related Experiment Video
Updated: Sep 20, 2025

A Training and Testing System for Performing Vascular Reconstruction In Vitro
Published on: October 26, 2019
Evaluating Pull-Out Strength of Barbed Suture In Vitro by Using Porcine Tissue and Polydimethylsiloxane (PDMS)
Wei Hong1, I-Cheng Chen1,2, Chen-Ying Su1
1Department of Chemical Engineering and Biotechnology, National Taipei University of Technology, No. 1, Sec. 3, Zhongxiao E. Rd., Taipei 10608, Taiwan.
Abstract:
Using barbed thread lifting for facial rejuvenation has become popular these days due to its minimally invasive procedures with reduced complications. However, only limited studies regarding its mechanical properties for face suspension were published. The aim of this study was to evaluate suture-holding ability regarding its facelift property, and different specimens were tested in order to establish an in vitro model. Fresh porcine tissue and the synthetic material polydimethylsiloxane (PDMS) were selected to simulate human skin for evaluating barbed suture pull-out strength by the universal material testing machine. The results showed that the pull-out strength of barbs between different porcine tissues varied without consistency. By contrast, PDMS (30:1) showed more consistent pull-out strength in each testing, and the average maximum load force was close to porcine tissue. Furthermore, after submerging barbed sutures in PBS for 0 days (T0), 7 days (T7) and 14 days (T14), a trend of decreased average maximum load force, displacement and force of 1.5 mm/2 mm/3 mm displacement could be detected by in vitro testing with PDMS (30:1). These results provide support for using PDMS (30:1) to evaluate suture pull-out strength and holding/lifting capacities in vitro to obtain consistent and objective information for evaluating substantial equivalence of devices. The established in vitro method could be used for the future development of barbed thread lifting technology.