Evaluation of a self-imaging SD-OCT system designed for remote home monitoring
Judy E Kim1, Oren Tomkins-Netzer2, Michael J Elman3
1The Eye Institute, Medical College of Wisconsin, 925 N. 87th street, Milwaukee, WI, 53226, USA. jekim@mcw.edu.
Purpose:
To compare identification rates of retinal fluid of the Notal Vision Home Optical Coherence Tomography (OCT) device (NVHO) when used by people with age-related macular degeneration (AMD) to those captured by a commercial OCT.
Methods:
Prospective, cross-sectional study where patients underwent commercial OCT imaging followed by self-imaging with either the NVHO 2.5 or the NVHO 3 in clinic setting. Outcomes included patients' ability to acquire analyzable OCT images with the NVHO and to compare those with commercial images.
Results:
Successful images were acquired with the NVHO 2.5 in 469/531 eyes (88%) in 264/290 subjects (91%) with the mean (SD) age of 78.8 (8.8); 153 (58%) were female with median visual acuity (VA) of 20/40. In the NVHO 3 cohort, 69 eyes of 45 subjects (93%) completed the self-imaging. Higher rates of successful imaging were found in eyes with VA ≥ 20/320. Positive percent agreement/negative percent agreement for detecting the presence of subretinal and/or intraretinal fluid when reviewing for fluid in three repeated volume scans were 97%/95%, respectively for the NVHO v3.
Conclusion:
Self-testing with the NVHO can produce high quality images suitable for fluid identification by human graders, suggesting the device may be able to complement standard-of-care clinical assessments and treatments.