A Cross-Sectional Study on the Agreement of Perfusion Indexes Measured on Different Fingers by a Portable Pulse
Sharada M Swain1, Manju Lata2, Sandeep Kumar2
1Department of Physiology, Hi-Tech Medical College and Hospital, Bhubaneswar, IND.
Abstract:
Background Pulse oximeters measure oxygen saturation, heart rate, and perfusion index (PI) by analyzing photoplethysmographic signals. PI is an indirect measure of peripheral perfusion expressed as a percentage of pulsatile signals to non-pulsatile signals. PI measured from different sites may show variation. PI may vary when measured on different fingers. In this study, we aimed to observe the variation of PI among different fingers of both hands. Methodology This cross-sectional, observational study was conducted using a convenience sample recruited from a tertiary care hospital in eastern India. PI was measured in apparently healthy adults in a sitting posture after a five-minute rest. The pulse oximeter probe was attached to each finger and readings were taken after one minute. The analysis of variance and intraclass correlation coefficient (ICC) were calculated to compare and find agreement among PI. Results Data from a total of 391 (229 [58.57%] male and 162 [41.43%] female) adult research participants with a mean age of 34.88 ± 10.65 years were analyzed. The PI was the highest on the middle finger in both hands. There was a significant difference among the PI measured on different fingers, F (9, 3900) = 15.49, p <0.0001. The ICC was 0.474, 0.368, and 0.635 for overall, right-hand, and left-hand fingers, respectively, which indicate poor (ICC < 0.5) to moderate (ICC = 0.5-0.75) levels of reliability. Conclusions The PI measured using consumer-grade pulse oximeters on different fingers may provide different readings. The highest PI reading is found on the middle finger. Clinicians and primary care physicians should consider the differences in measured PI among different fingers and should use the readings with caution for any diagnostic purposes.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Pulse Assessment Sites