Related Experiment Video
Updated: Sep 8, 2025

Assessment of Intestinal Transcytosis of Neonatal Escherichia coli Bacteremia Isolates
Published on: February 17, 2023
Contaminated Incubators: Source of a Multispecies Enterobacter Outbreak of Neonatal Sepsis
Enrique Hernandez-Alonso1, Nadège Bourgeois-Nicolaos1,2, Margaux Lepainteur2
1Institute of Integrative Biology of the Cell (I2BC), CNRS, CEA, Paris-Saclay University, Gif-sur-Yvette, France.
Abstract:
The genus Enterobacter includes species responsible for nosocomial outbreaks in fragile patients, especially in neonatal intensive care units (NICUs). Determining the primary source of infection is critical to outbreak management and patient outcomes. In this investigation, we report the management and control measures implemented during an Enterobacter outbreak of bloodstream infections in premature babies. The study was conducted in a French NICU over a 3-year period (2016 to 2018) and included 20 premature infants with bacteremia. The clinical and microbiological characteristics were identified, and whole-genome sequencing (WGS) was performed on bacteremia isolates. Initially, several outbreak containment strategies were carried out with no success. Next, outbreak investigation pinpointed the neonatal incubators as the primary reservoir and source of contamination in this outbreak. A new sampling methodology during "on" or "in use" conditions enabled its identification, which led to their replacement, thus resulting in the containment of the outbreak. WGS analysis showed a multiclonal outbreak. Some clones were identified in different isolation sources, including patients and neonatal incubators. In addition, microbiological results showed a multispecies outbreak with a high prevalence of Enterobacter bugandensis and Enterobacter xiangfangensis. We conclude that the NICU health care environment represents an important reservoir for Enterobacter transmission and infection. Finally, extracting samples from the neonatal incubator during active use conditions improves the recovery of bacteria from contaminated equipment. This method should be used more frequently to achieve better monitoring of the NICU for HAIs prevention. IMPORTANCE Neonatal incubators in the NICU can be an important reservoir of pathogens responsible for life-threatening outbreaks in neonatal patients. Traditional disinfection with antiseptics is not sufficient to eradicate the microorganisms that can persist for long periods in the different reservoirs. Identification and elimination of the reservoirs are crucial for outbreak prevention and control. In our investigation, using a new strategy of microbiological screening of neonatal incubators, we demonstrated that these were the primary source of contamination. After their replacement, the outbreak was controlled. This new methodology was effective in containing this outbreak and could be a viable alternative for infection prevention and control in outbreak situations involving incubators as a reservoir.
Insights
Neonatal incubators were the source of an Enterobacter outbreak in a French neonatal intensive care unit (NICU). Replacing contaminated incubators and using new sampling methods controlled the bloodstream infections in premature infants.
Area of Science:
- Medical Microbiology
- Infectious Disease Epidemiology
Background:
- Enterobacter species cause nosocomial outbreaks, particularly in neonatal intensive care units (NICUs), posing significant risks to vulnerable premature infants.
- Effective outbreak management hinges on identifying the primary source of infection to implement targeted control measures.
Purpose of the Study:
- To investigate and control an Enterobacter bloodstream infection outbreak in premature infants within a French NICU.
- To identify the reservoir and source of contamination contributing to the outbreak.
Main Methods:
- Conducted a 3-year study (2016-2018) involving 20 premature infants with bacteremia.
- Performed whole-genome sequencing (WGS) on clinical isolates and employed a novel sampling methodology for neonatal incubators during active use.
- Implemented and evaluated outbreak containment strategies, including incubator replacement.
Main Results:
- Initial containment strategies were unsuccessful.
- Neonatal incubators were identified as the primary reservoir and source of Enterobacter contamination using the new sampling method.
- WGS revealed a multiclonal and multispecies outbreak, with Enterobacter bugandensis and Enterobacter xiangfangensis being prevalent.
- Replacement of contaminated incubators successfully contained the outbreak.
Conclusions:
- The NICU healthcare environment, specifically neonatal incubators, serves as a significant reservoir for Enterobacter transmission and infection.
- A novel sampling methodology for incubators during active use enhances bacterial recovery and is crucial for effective HAI prevention and monitoring.
- Identifying and eliminating contaminated reservoirs, like incubators, is vital for controlling outbreaks in NICUs.