Establishing Reference Range and Assessing Precision Performance of the TEG6s System; Verification of TEG6s to
Background:
Thromboelastography (TEG) provides assessment of global coagulation. The TEG6s (Haemonetics Corporation) is a newly developed cartridge-based system, fully automated and the first true point-of-care TEG. In this study, we evaluated the precision and established the reference intervals (RIs) for TEG6s.
Methods:
TEG assays were performed on the blood of healthy donors to determine RIs and on the QC materials for precision testing. The study design was developed in accordance with Clinical and Laboratory Standards Institute Guidelines.
Results:
TEG6s precision testing yielded low variability except R, due to a low value for the mean. The newly established RIs of R, MA, and LY30 were similar to the manufacturer's RIs. Some were different, showing short K and increased α.
Conclusions:
This study has shown that TEG6s has high precision and each institution should verify the manufacturer's RIs before adopting TEG6s and establish RIs if necessary.
More Related Videos
13:24Integration of Wet and Dry Bench Processes Optimizes Targeted Next-generation Sequencing of Low-quality and Low-quantity Tumor Biopsies
Published on: April 11, 2016
08:52Improving Reproducibility to Meet Minimal Information for Studies of Extracellular Vesicles 2018 Guidelines in Nanoparticle Tracking Analysis
Published on: November 17, 2021
Related Concept Videos
Data Validation
Key parameters for method validation include:
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...