Dosimetry for Cell Irradiation using Orthovoltage 40-300 kV X-Ray Facilities
Published on: February 20, 2021
Dosimetry audits in Taiwan radiotherapy departments
An-Cheng Shiau, Shih-Ming Hsu1, Pei-Yun Huang2
1Department of Biomedical Imaging and Radiological Sciences, National Yang-Ming University, Taipei, Taiwan.
Objectives:
This study examines the practice of the regulation of Standards for Medical Exposure Quality Assurance (SMEQA) in Taiwan based on on-site quality audit for radiation therapy systems from 2016 to 2019.
Methods:
81 radiation therapy departments, 141 linacs, 9 γ knife systems, 34 high dose rate brachytherapy systems, 20 Tomotherapys, and 6 Cyberknives were audited yearly. Data collection and analysis for each institute's documents including QA procedure, ion chamber and electrometer calibration reports, and a questionnaire relating to machine type and staffing, were requested first and reviewed by auditors. On-site SMEQA core item measurements, including beam output, beam profile and energy constancy for external beam therapy systems, and the source strength, positioning, and timer accuracy for brachytherapy systems were audited second. More than 300 photon beams and more than 400 electron beams were measured each year.
Results:
There were approximately 8.9 radiotherapy units per million population, and 1.2 medical physicists per unit in Taiwan. For the output measurements, more than 78 and 75% of the photon beams and electron beams, respectively, from linacs were with deviations within ±1.0%. Photon beams have lower beam quality measurement deviations than electron beams. Including in-plane and cross-plane measurements, more than 90 and 85% photon and electron beams, respectively, were with flatness consistency within 1.0%. All audit measurements were within the SMEQA acceptance criteria.
Conclusions:
According to SMEQA regulations on-site QA audits were successfully carried out from 2016 to 2019 for all Taiwan radiotherapy units. The measurement results showed high quality machine performance in Taiwan.
Advances In Knowledge:
Dosimetry audits with directly acquired measurement readings have lower uncertainties; allow immediate feedback, discussion, and adjustment in a timely manner. In addition to regulation system establishment and education and training implementation, the machine quality is closely related to machine maintenance implementation.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET