Influence of eye tilt on corneal densitometry
Alejandra Consejo1, Marta Jiménez-García2,3, Jos J Rozema2,3
1Department of Applied Physics, University of Zaragoza, Zaragoza, Spain.
Purpose:
To investigate whether Pentacam densitometry readings are affected by corneal tilt.
Methods:
In a prospective study, the right eyes of 86 healthy participants aged 42.8 ± 20.0 years (range 18-79 years) were imaged using Scheimpflug tomography. Elevation maps were exported to calculate corneal tilt using custom-made software, and densitometry readings were acquired directly from the corneal densitometry analysis add-on to the standard software Oculus Pentacam HR. Simple mediation analysis was applied to study age as a confounding factor in the correlation between corneal tilt and corneal densitometry.
Results:
Corneal tilt and corneal densitometry are not independent from one another because age is significantly correlated with both corneal tilt (r = 0.50, p < 0.001) and corneal densitometry (r = 0.91, p < 0.001). Only 3.8% of the correlation between tilt and densitometry operates directly, while the remaining 96.2% depends on age.
Conclusions:
Corneal tilt plays a role in corneal densitometry readings, even though the interaction is strongly influenced by age. Age is a well-known factor in densitometry readings that should be taken into consideration when interpreting Scheimpflug densitometry.