Reliability and Validity of Standing Lateral Radiograph Method for Measuring Acetabular Component Version: A Modified
Wenhui Zhang1, Jie Xu1, Deng Li1
1Sun Yat-sen Memorial Hospital, Sun Yat-sen University, Guangzhou, China.
Orthopaedic Surgery
|June 17, 2022
Summary
Standing lateral radiography is a reliable and valid method for measuring acetabular component version after total hip arthroplasty (THA). This technique accurately assesses cup versions, showing excellent consistency with CT scans and minimal differences in measurements across various X-ray incidence angles.
Area of Science:
- Orthopedic Surgery
- Radiology
- Medical Imaging
Background:
- Accurate measurement of acetabular component version is crucial for successful total hip arthroplasty (THA) outcomes.
- Traditional methods like cross-table lateral (CL) radiography can be prone to measurement errors due to pelvic tilt.
- Standing lateral (SL) radiography offers a potential alternative for more accurate cup version assessment.
Purpose of the Study:
- To evaluate the impact of X-ray incidence angle on cup version measurements.
- To determine the reliability and validity of standing lateral (SL) radiography for measuring cup versions.
- To compare SL radiography measurements with computed tomography (CT) scans.
Main Methods:
- 3D simulation analysis was used to investigate cup versions under varying X-ray incidence angles.
- Retrospective analysis of 93 patients who underwent primary THA, comparing SL radiographs with CT scans.
- Cup versions were measured by two orthopedic physicians, with intra- and inter-observer reliabilities assessed using intra-class correlation coefficient (ICC).
Main Results:
- No significant differences in cup version measurements were found across different X-ray incidence angles (P = 0.663).
- Excellent intra- and inter-observer reliabilities were achieved (ICC > 0.95).
- SL radiograph measurements showed strong correlation with CT scans (r = 0.853, P < 0.001), with a mean difference of -0.49°.
Conclusions:
- Standing lateral (SL) radiography provides cup version measurements comparable to CT scans.
- The SL radiography method demonstrates high reliability and validity for assessing acetabular component version post-THA.
- SL radiography is a dependable tool for postoperative evaluation of cup version in total hip arthroplasty.
Keywords:
Acetabular component versionCross-table lateral radiographHip dislocationStanding lateral radiographTotal hip arthroplasty