Atomic-resolution STEM image denoising by total variation regularization

Kazuaki Kawahara1, Ryo Ishikawa1, Shun Sasano1

  • 1Institute of Engineering Innovation, The University of Tokyo, Bunkyo, Tokyo 113-8656, Japan.

Related Concept Videos

Deconvolution01:20

Deconvolution

Deconvolution, also known as inverse filtering, is the process of extracting the impulse response from known input and output signals. This technique is vital in scenarios where the system's characteristics are unknown, and they must be inferred from the observable signals.
Deconvolution involves several mathematical techniques to derive the impulse response. One common approach is polynomial division. In this method, the input and output sequences are treated as coefficients of...
246
Super-resolution Fluorescence Microscopy01:37

Super-resolution Fluorescence Microscopy

Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been...
7.5K
Electron Microscope Tomography and Single-particle Reconstruction01:07

Electron Microscope Tomography and Single-particle Reconstruction

Transmission electron microscopy (TEM) can be used to determine the 3D structure of biological samples with the help of techniques such as electron microscope tomography and single-particle reconstruction. While single-particle reconstruction can examine macromolecules and macromolecular complexes in vitro conditions only, tomography permits the study of cell components or small cells in vivo.
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
2.5K
Root Mean Square00:57

Root Mean Square

If in an experiment, data values have a probability of being both positive and negative, neither the arithmetic mean, the geometric mean, nor the harmonic mean can be used to calculate the central tendency of the data set. In particular, if the positive and negative values are equally likely, the arithmetic mean is close to zero.
For example, consider the velocity of gas molecules in a container. The gas molecules are moving in different directions, which might impart positive and negative...
3.3K
Residuals and Least-Squares Property01:11

Residuals and Least-Squares Property

The vertical distance between the actual value of y and the estimated value of y. In other words, it measures the vertical distance between the actual data point and the predicted point on the line
If the observed data point lies above the line, the residual is positive, and the line underestimates the actual data value for y. If the observed data point lies below the line, the residual is negative, and the line overestimates the actual data value for y.
The process of fitting the best-fit...
7.8K
Difference from Background: Limit of Detection01:05

Difference from Background: Limit of Detection

The limit of detection (LOD) is the smallest amount of analyte that can be distinguished from the background noise. The LOD value corresponds to the concentration at which the analyte signal is three times larger than the standard deviation of the blank signal. Below this value, the analyte signal cannot be differentiated from the background noise. It is calculated by dividing the calibration slope by 3 times the standard deviation of the blank signals.
The LOD indicates the presence or absence...
7.0K