Contrast Enhanced Vessel Imaging using MicroCT
Published on: January 27, 2011
Effects of contrast enhancement boost postprocessing technique in combination with different reconstruction
Jia Xu1, Shitian Wang1, Xuan Wang1
1Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Objectives:
To investigate whether contrast-enhancement-boost (CE-boost) in combination with hybrid iterative reconstruction (Hybrid IR, also named HIR [AIDR 3D, adaptive iterative dose reduction three dimensional]) and model-based iterative reconstruction (MBIR [FIRST, forward projected model-based IR solution]) algorithms can improve the image quality of abdominal CT angiography (CTA).
Methods:
This retrospective study included 50 patients who underwent abdominal CTA. Both arterial and portal phases were reconstructed using three different algorithms [filtered-back projection (FBP), AIDR 3D, and FIRST] separately. CE-boost was performed additionally to generate AIDR 3D-boost and FIRST-boost images. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of the arteries and portal system were compared among the five datasets (FBP, AIDR 3D, FIRST, AIDR 3D-boost, FIRST-boost). In subjective analyses, two radiologists independently ordered images (5, best; 1, worst) based on the visual image quality of distal arterial or portal venous branches. The Friedman and the Dunn-Bonferroni post-hoc tests were used for statistical analysis.
Results:
FIRST-boost arterial and portal images had the lowest noise compared with FBP, AIDR 3D, FIRST, and AIDR 3D-boost images (all P < 0.05), and significantly higher SNR and CNR than FBP, AIDR 3D, and FIRST images (all P < 0.05). AIDR 3D-boost images showed lower noise, and higher SNR and CNR than FBP and AIDR 3D images (all P < 0.05). FIRST-boost images had higher subjective grading scores than FBP, AIDR 3D, and AIDR 3D-boost images (all P < 0.05).
Conclusion:
The postprocessing technique CE-boost can improve the image quality of abdominal CTA images. MBIR in combination with CE-boost (FIRST-boost) images had the best image quality compared with the other four image datasets.
More Related Videos
06:02Anatomical Reconstructions of the Human Cardiac Venous System using Contrast-computed Tomography of Perfusion-fixed Specimens
Published on: April 18, 2013
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...