A rapid, objective method for the detection of lymphocytotoxic antibodies using flow cytometry
Journal of Immunological Methods
|May 4, 1987
Abstract:
Whilst several centres have reported lymphocytotoxic antibody detection using single and dual fluorescent stains with analysis of the fluorescence emitted from the cell population present in a well of a multiwell plate, problems are encountered with cell concentration and light emission overlap. A method we have developed using flow cytometry produces similar values of percentage cell death using single or double staining techniques (correlation coefficient = 0.9896). This method is not influenced by slight variation in cell number or light emission overlap. The effect of introducing red cell impurities into the normal lymphocyte preparation is described.