Related Experiment Video
Updated: Sep 7, 2025

Ex Vivo Porcine Experimental Model for Studying and Teaching Lung Mechanics
Published on: April 19, 2024
Comparison of nasal intermittent positive pressure ventilation and bubble CPAP with an in-line high-frequency
David M Rub1,2, Emidio M Sivieri1,3, Soraya Abbasi1,2,3
1Division of Neonatology, Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA.
Introduction:
Noninvasive ventilation has become a staple in the care of premature infants. However, failure rates continue to be high in this population. Modifications to noninvasive support, such as nasal intermittent positive pressure ventilation (NIPPV), are used clinically to reduce such failure. Previous in vitro studies have shown improved CO2 clearance when superimposing high-frequency oscillations onto bubble continuous positive airway pressure (BCPAP).
Objective:
To compare the CO2 clearance of NIPPV to BCPAP with an in-line high-frequency interrupter (HFI) in a premature infant lung model.
Methods:
A premature infant lung model was connected to either a Dräger VN500 for delivery of NIPPV or a BCPAP device with superimposed high-frequency oscillations generated by an in-line HFI. Change in end-tidal CO2 (ETCO2 ) and mean airway pressure at the simulated trachea were measured and compared for both noninvasive modalities.
Results:
Superimposing HF oscillations onto BCPAP with an in-line HFI resulted in improved CO2 clearance relative to BCPAP alone for all tested oscillation frequencies at all CPAP levels (p < 0.001). NIPPV also resulted in improved CO2 clearance relative to nasal CPAP (NCPAP) alone (p < 0.001). Among the tested settings, BCPAP with an in-line HFI resulted in decreased ETCO2 relative to BCPAP ranging from -14% to -36%, while NIPPV resulted in decreased ETCO2 relative to NCPAP ranging from -2% to -12%.
Conclusion:
Superimposing high-frequency oscillations onto BCPAP using a novel in-line HFI was found to be more effective at clearing CO2 than NIPPV in a premature infant lung model.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Cardiopulmonary Resuscitation II: ACLS Airway Management