Hypoglossal Nerve Stimulator Explantation: A Case Series
Anuj U Patel1, Pratyusha Yalamanchi1, Nithin S Peddireddy1
1Department of Otolaryngology-Head and Neck Surgery, University of Michigan, Ann Arbor, Michigan, USA.
OTO Open
|June 23, 2022
Summary
Hypoglossal nerve stimulation (HGNS) explantation is uncommon but necessary for reasons like MRI incompatibility, infection, or cosmetic concerns. Consider future medical needs before HGNS implantation for sleep apnea.
Area of Science:
- Neurosurgery
- Sleep Medicine
- Medical Device Technology
Background:
- Hypoglossal nerve stimulation (HGNS) is an effective treatment for moderate to severe obstructive sleep apnea (OSA) in patients unresponsive to positive airway pressure (PAP) therapy.
- Adverse events leading to HGNS explantation are infrequent and not well-documented.
- This study presents the first case series detailing reasons for HGNS system explantation.
Purpose of the Study:
- To describe the incidence and reasons for explantation of the Inspire HGNS system.
- To identify key factors influencing the decision for HGNS explantation.
- To inform future patient selection and management strategies for HGNS therapy.
Main Methods:
- Retrospective review of five patients who underwent explantation of the Inspire HGNS system.
- Analysis of explantation indications, operative times, and patient outcomes.
- Case series methodology to document clinical experiences.
Main Results:
- Five patients underwent HGNS explantation.
- Primary reasons for explantation included magnetic resonance imaging (MRI) incompatibility (3 patients), poor cosmesis (1 patient), and surgical site infection (1 patient).
- Average operative explant time was 163 minutes.
Conclusions:
- MRI incompatibility, poor cosmesis, and device-related infection are significant reasons for HGNS explantation.
- Potential future requirements for MRI or chest wall surgery should be carefully evaluated during the pre-implantation assessment for HGNS.
- This case series highlights important considerations for the long-term management and potential removal of HGNS devices.