Related Experiment Video
Updated: Sep 7, 2025

Characterization of the Isolated, Ventilated, and Instrumented Mouse Lung Perfused with Pulsatile Flow
Published on: April 29, 2011
Fraction of Inspired Oxygen With Low-Flow Versus High-Flow Devices: A Simulation Study
Yuki Kojima1, Ryozo Sendo2, Naoko Okayama3
1Department of Anesthesiology, Asahi General Hospital, Asahi, JPN.
Purpose:
The fraction of inspired oxygen while administering oxygen to patients must be measured as it represents the alveolar oxygen concentration, which is important from a respiratory physiology viewpoint. Therefore, the purpose of this study was to compare the fractions of inspired oxygen obtained through different oxygen delivery devices.
Methods:
A simulation model of spontaneous respiration was used. The fractions of inspired oxygen obtained through low- and high-flow nasal cannulas and a simple oxygen mask were measured. The fraction of inspired air was measured every second for 30 s after 120 s of oxygen administration. This was measured three times under each condition.
Results:
With a low-flow nasal cannula, airflow reduced both the intratracheal fraction of inspired oxygen and extraoral oxygen concentration, indicating that exhalatory respiration occurred during rebreathing and may be involved in increasing the intratracheal fraction of inspired oxygen.
Conclusion:
Oxygen administration during expiratory flow may lead to an increased oxygen concentration in the anatomical dead space, which may be involved in the increase in the fraction of inspired oxygen. With a high-flow nasal cannula, a high fraction of inspired oxygen can be achieved even at a flow rate of 10 L/min. When determining the optimum amount of oxygen, it is necessary to set an appropriate flow rate for patients and specific conditions without being bound by the fraction of inspired oxygen values alone. It might be difficult to estimate the fraction of inspired oxygen while using a low-flow nasal cannula and simple oxygen mask in clinical situations.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...