Introduction:
The Hannover Coupler version 2 (HC2) was designed to (1) adapt the coupler geometry to the round window (RW) niche (2) to stabilize the floating mass transducer, and (3) to control static coupling forces to the RW. First audiological outcomes with a custom-made HC2 are reported here.
Material And Methods:
Ten patients were enrolled in our site-initiated, prospective study. To assess audiological outcomes up to 6 months, preoperative and postoperative hearing thresholds, word recognition score (WRS) at 65 dB SPL and the speech recognition threshold in quiet and noise were performed. The effective gain (EG) and the coupling efficiency were calculated.
Results:
One revision surgery had to be performed during the study period and a significant, but clinically not relevant bone conduction thresholds change was observed at 4 and 6 kHz at 6-month follow-up. At 6 months, the median WRS (n = 10) improved significantly from 0% to 80%. The median speech reception threshold in noise improved significantly from 11.6 to -2.4 dB SNR, and in quiet significantly from 79.6 to 44.4 dB SPL. The average EG of -1.3 dB indicated a closure of the air bone gap. The determined average coupling efficiency of 23.3 dB was within the acceptance range suggested by the manufacturer.
Conclusion:
For patients with mixed hearing loss and multiple ear surgeries, the HC2 provided good and stable speech recognition results exceeding published results of RW coupling without a coupler or coupling with the RW soft coupler.
Related Concept Videos
Anatomy of the Ear
Wheatstone Bridge
Thus, for accurate resistance measurements, a...
The Auditory Ossicles
The aptly named stapes look very much like a stirrup. The three ossicles are unique to mammals, and each plays a role in...
The Cochlea
Hearing
Auditory Pathway
When viewed cross-sectionally, the cochlea reveals the scala vestibuli and scala tympani flanking...