Influence of the Fluoroscopy Setting towards the Patient When Identifying the MPFL Insertion Point
Alexander Korthaus1, Tobias Dust1, Markus Berninger1
1Department of Trauma and Orthopaedic Surgery, University Medical Center Hamburg-Eppendorf, 20246 Hamburg, Germany.
Diagnostics (Basel, Switzerland)
|June 24, 2022
Summary
Femoral tunnel malposition in medial patellofemoral ligament (MPFL) reconstruction causes graft issues. Intraoperative fluoroscopy positioning significantly impacts MPFL tunnel placement, necessitating careful attention during surgery.
Area of Science:
- Orthopedic Surgery
- Biomedical Engineering
- Radiology
Background:
- Femoral tunnel malposition in medial patellofemoral ligament (MPFL) reconstruction is a leading cause of revision surgery.
- Incorrect tunnel placement can alter graft length and increase patellofemoral joint pressure.
Purpose of the Study:
- To investigate if intraoperative fluoroscopy positioning affects femoral tunnel accuracy in MPFL reconstruction.
- To evaluate the impact of image receptor distance and direction on determining the correct femoral origin of the MPFL.
Main Methods:
- 3D-printed femurs from CT scans were used to simulate MPFL reconstruction.
- True-lateral fluoroscopies were performed at different image receptor distances (25 cm and 5 cm) and directions (lateral-to-medial and medial-to-lateral).
- MPFL femoral origin points were determined using the Schöttle et al. method and compared across different fluoroscopic setups.
Main Results:
- Determining the MPFL origin in a 5 cm medial-to-lateral view resulted in a significant distal and posterior shift (5.3 mm) when checked in a 25 cm lateral-to-medial view.
- Conversely, determining the origin in the 25 cm lateral-to-medial view and rechecking in the 5 cm medial-to-lateral view caused an anterior and proximal shift (4.8 mm).
- Variations in femur positioning (distal, proximal, superior, inferior) did not significantly influence the results.
Conclusions:
- Fluoroscopy positioning is critical for accurate femoral tunnel placement in MPFL reconstruction.
- Surgeons must consider the fluoroscopy's position relative to the image receptor, not just a true-lateral view, to avoid MPFL graft malposition.
- Attention to fluoroscopic angulation is crucial for successful MPFL reconstruction outcomes.