Experience of Combined Procedure during Percutaneous LAA Closure
Related Experiment Video
Updated: Sep 6, 2025

Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Guillaume Domain1, Nicolas Dognin1, Gilles O'Hara1
1Institut Universitaire de Cardiologie et de Pneumologie de Québec, Quebec City, QC G1V 4G5, Canada.
Introduction:
Percutaneous left atrial appendage closure (LAAC) is an alternative to oral anticoagulants (OAC) in patients with non-valvular atrial fibrillation (AF) and contraindication to long-term OAC. Combined strategy with percutaneous LAAC at the same time of other cardiac structural or electrophysiological procedures has emerged as an alternative to a staged strategy.
Aim:
To describe our experience with combined LAAC procedures using Watchman™ devices.
Methods:
All patients with combined LAAC procedures using Watchman™ (WN) devices performed from 2016 to 2021 were included. The primary safety endpoint was a composite of periprocedural complications and adverse events during the follow-up. The primary efficacy endpoint included strokes, systemic embolisms, major bleeding and cardiovascular death.
Results:
From 2016, among 160 patients who underwent LAAC using WN devices, 19 underwent a combined strategy: 7 transcatheter edge-to-edge mitral valve repair (TEMVR) (37%), 6 typical atrial flutter ablation (31%), 2 leadless pacemaker (LP) implantation (10%) and 4 AF ablation (22%). The WN device was successfully implanted in 98% and 100% of cases for single and combined LAAC procedures, respectively (p = 0.63). Median follow-up was 13 months (IQR 25/75 3/24). Device-related complications occurred in 6 out of 141 patients (4%) who underwent single LAAC and in no (0/19) patient in the combined LAAC procedure (p = ns). The procedural-related complications did not differ significantly between groups (5% vs. 10%, respectively, in the single and combined group, p = 0.1).
Conclusion:
Combined procedure using the Watchman™ devices and one other structural or electrophysiological procedure appears safe and effective. Larger series are needed to confirm these results.