Related Experiment Video
Updated: Sep 6, 2025

10:19
3D Kinematic Gait Analysis for Preclinical Studies in Rodents
Published on: August 3, 2019
10.8K
Evaluation of lower extremity gait analysis using Kinect V2® tracking system.
Takuya Usami1, Kazuki Nishida2, Hirotaka Iguchi3
1Department of Orthopedic Surgery, Nagoya City University Graduate School of Medical Sciences, Nagoya, Aichi 467-8601, Japan - Nagoya City University East Medical Center, Nagoya, Aichi 464-8547, Japan.
SICOT-J
|June 24, 2022
Summary
Microsoft Kinect V2 is a reliable device for human gait analysis, accurately measuring gait velocity, step length, and joint angles. This technology offers a cost-effective alternative for gait evaluation.
Area of Science:
- Biomechanics
- Human Movement Analysis
- Wearable Technology
Background:
- Microsoft Kinect V2 is a depth-sensing peripheral device.
- It captures depth, posture, and skeleton data.
- Its utility for human gait analysis requires investigation.
Purpose of the Study:
- To evaluate the reliability of the Microsoft Kinect V2 for human gait analysis.
- To compare Kinect V2 gait parameters with a marker-based motion capture system.
Main Methods:
- Ten healthy volunteers performed 20 walking trials each.
- Gait was recorded simultaneously using Kinect V2 and an infrared marker-based system.
- Pearson's correlation and meta-analysis assessed reliability of gait velocity, cycle time, step length, joint angles, ground contact time, and ankle velocity.
Main Results:
- Strong reliability (r=0.92) was found for gait velocity and step length (r=0.81).
- Moderate reliability was observed for gait cycle time (r=0.65) and minimum hip flexion angle (r=0.68).
- Fair to poor reliability was noted for other parameters, with similar hip and knee angle trajectories between systems.
Conclusions:
- Microsoft Kinect V2 demonstrates strong reliability for key gait parameters like velocity, cycle time, and step length.
- It is a potentially valuable and accessible tool for gait analysis.
- Further research may refine its application for specific joint angle measurements.