Implementation and Validation of a Novel Noninvasive blood Pressure Measurement Algorithm Based on Oscillometric
Sahar Karimian1, Mohammadreza Yazdchi2, Reza Hajian3
1Department of Biomedical Engineering, University of Isfahan, Isfahan, Iran.
Background:
One of the most prevalent methods in noninvasive blood pressure (BP) measurement with cuff is oscillometric, which has two different types of deflation, including linear and step deflation. With this approach, in addition to designing a novel algorithm by the step deflation method, a sample of its module was constructed and validated during clinical tests in different hospitals.
Method:
In this study, by controlling the valve, the pressure would be deflated through optimized steps. By real-time processing on the obtained signal from the pressure sensor, pulses in each step would be extracted. After that, in offline mode, mean arterial pressure is estimated based on curve fitting.
Result:
A BP simulator, various modules, and an auditory method were used to validate the algorithm and its results. During clinical tests, 80 people (men and women), 11 dialysis patients, and 69 non-dialysis (healthy or with other diseases) in the age range of 17-85 years participated.
Conclusion:
The obtained results compared with the BP simulator are in the standard range according to the international medical standards of the British Hypertension Society (BHS) and the US Association for the Advancement of Medical Instrumentation (AAMI), which are the global standard of comparison in this field.
More Related Videos
14:09Tilt Testing with Combined Lower Body Negative Pressure: a "Gold Standard" for Measuring Orthostatic Tolerance
Published on: March 21, 2013
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure