Radiation Dose Reduction in Contrast-Enhanced Abdominal CT: Comparison of Photon-Counting Detector CT with 2nd
Robin Wrazidlo1, Lukas Walder1, Arne Estler1
1Department of Diagnostic and Interventional Radiology, Eberhard-Karls-University, 72076 Tübingen, Germany (R.W., L.W., A.E., K.N., M.H., F.H.); Siemens Healthcare GmbH, 91052 Erlangen, Germany (R.G., B.S., S.F.); NYU Grossman School of Medicine, Department of Radiology, New York, NY 10016, USA (J.F.).
Rational And Objectives:
Comparison of radiation dose and image quality in routine abdominal and pelvic contrast-enhanced computed tomography (CECT) between a photon-counting detector CT (PCD-CT) and a dual energy dual source CT (DSCT).
Materials And Methods:
70 oncologic patients (mean age 66 ± 12 years, 29 females) were prospectively enrolled between November 2021 and February 2022. Abdominal CECT were clinically indicated and performed first on a 2nd-generation DSCT and at follow-up on a 1st-generation dual-source PCD-CT. The same contrast media (Imeron 350, Bracco imaging) and pump protocol was used for both scans. For both scanners, polychromatic images were reconstructed with 3mm slice thickness and comparable kernel (I30f[DSCT] and Br40f[PCD-CT]); for PCD-CT data from all counted events above the lowest energy threshold at 20 keV ("T3D") were used. Results were compared in terms of radiation dose metrics of CT dose index (CTDIvol), dose length product (DLP) and size-specific dose estimation (SSDE), objective and subjective measurements of image quality were scored by two emergency radiologists including lesion conspicuity.
Results:
Median time interval between the scans was 4 months (IQR: 3-6). CNRvessel and SNRvessel of T3D reconstructions from PCD-CT were significantly higher than those of DSCT (all, p < 0.05). Qualitative image noise analysis from PCD-CT and DSCT yielded a mean of 4 each. Lesion conspicuity was rated significantly higher in PCD-CT (Q3 strength) compared to DSCT images. CTDI, DLP and SSDE mean values for PCD-CT and DSCT were 7.98 ± 2.56 mGy vs. 14.11 ± 2.92 mGy, 393.13 ± 153.55 mGy*cm vs. 693.61 ± 185.76 mGy*cm and 9.98 ± 2.41 vs. 14.63 ± 1.63, respectively, translating to a dose reduction of around 32% (SSDE).
Conclusion:
PCD-CT enables oncologic abdominal CT with a significantly reduced dose while keeping image quality similar to 2nd-generation DSCT.
More Related Videos
08:30X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
X-ray Imaging
Imaging Studies for Cardiovascular System V: CT