Recall devices: Defects at initial visit and during remote monitoring supported follow-up
Maciej Grymuza1, Lidia Chmielewska-Michalak1, Agnieszka Katarzyńska-Szymańska1
11st Department of Cardiology, Poznan University of Medical Sciences, Poznań, Poland.
Background:
In March 2021, Biotronik informed about the risk of premature battery depletion in a group of implantable cardioverter-defibrillators. Following the manufacturers' recommendation, our center executed a recall and introduced remote monitoring (RM) in patients with susceptible devices. This study reports the rate of premature battery depletion in our center and events found in RM-supported follow-up.
Methods:
Single-center observational study.
Results:
Out of the 206 susceptible implanted devices, 125 patients appeared for the visit and RM was introduced in 107 (83%) patients. Until the visit, three (2.4%) devices required replacement due to battery depletion, and a further three (2.4%) devices had unexpected battery depletion during follow-up.
Conclusions:
The recalled devices had a higher rate of battery exhaustion than expected, while other device or lead defects were less common.