Related Experiment Video
Updated: Sep 6, 2025

Microsurgical Clip Obliteration of Middle Cerebral Aneurysm Using Intraoperative Flow Assessment
Published on: September 25, 2009
Flow Diverting Stents in Cerebral Small Caliber Vessels (< 2 mm) for Aneurysm Treatment : A Three Center
Sophia Hohenstatt1, Sergio L Vinci2, Dominik F Vollherbst1
1Department of Neuroradiology, Heidelberg University Hospital, Heidelberg, Germany.
Purpose:
The off-label use of flow diverting stents (FDS) for treating cerebral aneurysms in small distal vessels is increasing in clinical practice with encouraging results; however, data directly addressing the parent vessel size are still scarce. Our aim was to evaluate the safety and efficacy of FDS placement in anterior and posterior circulation aneurysms with parent arteries ≤ 2 mm in a real-world representative setting.
Methods:
We retrospectively reviewed patients treated with FDS at the three participating university hospitals between 2009 and 2021. The inclusion criteria were the placement of at least one FDS in a parent vessel with a maximum diameter of 2 mm or less. The primary clinical safety endpoint was the absence of death, major or minor symptomatic stroke, transient ischemic attack and procedure-related intracranial hemorrhage. Clinical outcome was assessed using the modified Rankin Scale (mRS) score at the follow-up visit at 1 year. The primary efficacy endpoint was defined as complete and near-complete occlusion at the 1‑year follow-up.
Results:
We identified 55 patients harboring 56 aneurysms. The primary clinical safety endpoint was obtained in 93% of cases. The hemorrhagic and thromboembolic complication rates were 4% and 9%, respectively. No patient died or had a relevant discrepancy (> 1 point) at the pre-mRS and post-mRS. The primary efficacy endpoint was reached in 80% of cases.
Conclusion:
The use of FDS in vessels < 2 mm is technically feasible with good aneurysm occlusion rates and an acceptable safety profile. Nevertheless, it is essential to be aware of the of the main complications associated when operating in small diameter vessels.
More Related Videos