Evaluation of outflow temperatures generated by a dry heat fluid warmer at low fluid rates
Hannah L Badham1, Louise Clark1, Sarah Gibson1
1Davies Veterinary Specialists, Higham Gobion, Hertfordshire, UK.
Objective:
To evaluate the output temperature of a dry heat fluid warmer at fluid rates typically used in small animal veterinary patients.
Study Design:
Prospective in vitro study.
Animals:
None.
Methods:
Ambient temperature lactated Ringer's (17.9-18.8 °C) was delivered via a dry heat fluid warmer. A temperature probe was used to measure fluid outflow temperature from the compatible giving set at 5, 10, 20, 50, 70, 80, 100, 200, 300, 400 and 500 mL hour-1. Outflow fluid temperature at plateau (two consecutive readings within 0.1 °C) was compared with baseline fluid temperature (fluid warmer turned off) to calculate temperature changes at each rate. Kruskal-Wallis test was used to compare changes in temperature and time to plateau temperature. Dunn's post hoc test was used to test for significant differences in temperature compared to 5 mL hour-1; p value < 0.05.
Results:
Median plateau outflow temperature increased as fluid rate increased, with temperatures of 18.5, 18.6, 18.7, 18.8, 19.4, 19.4, 21.5, 25.3, 28.5, 30.7 and 32.6 °C, at flow rates of 5, 10, 20, 50, 70, 80, 100, 200, 300, 400 and 500 mL hour-1. Fluid rates > 100 mL hour-1 showed significant increases from baseline (p = 0.021) There was no difference in temperature change from baseline at fluid rates < 100 mL hour-1 (p > 0.05). Compared to plateau temperature at 5 mL hour-1, there was a statistical difference in plateau temperature above 100 mL hour-1(p = 0.0207). Maximum outflow plateau temperature was 32.6 °C at 500 mL hour-1.
Conclusions And Clinical Relevance:
A dry heat fluid warmer has significantly decreased efficacy at low fluid rates, with no statistically significant increase in fluid temperature at fluid rates below 100 mL hour-1 at the end of a compatible fluid line. Inline dry heat fluid warmers are ineffective at fluid rates below 100 mL hour-1.
More Related Videos
Related Concept Videos
Temperature and Thermal Equilibrium
The concept of temperature has evolved from the common concepts of hot and cold. The scientific definition of temperature explains more than just our sense of hot and cold. Temperature is operationally defined as the quantity measured with a thermometer. Furthermore, temperature is...
Methods of reducing fever
Pharmacological Methods of Reducing Fever:
Decreased Body Temperature
The Carnot Cycle
What could be the theoretical limit to the efficiency of a heat engine? The...
Constant Pressure Calorimetry
Mechanisms of Heat Transfer II