Related Experiment Video
Updated: Sep 6, 2025

Evaluation of a Novel Laser-assisted Coronary Anastomotic Connector - the Trinity Clip - in a Porcine Off-pump Bypass Model
Published on: November 24, 2014
Laboratory "In-vitro" Evaluation of the Parallel Stent Graft Association for the Iliac Sandwich Technique
Rafael Narciso Franklin1,2, Jorge Rufino Ribas Timi3, Guilherme Baumgardt4
1Coris Cirurgia Vascular e Endovascular, Rua Menino Deus, 63 sala 504, Baia Sul Medical Center, Centro, Florianópolis, SC, 88020-210, Brazil. rafael.n.franklin@coris.med.br.
Objectives:
The Iliac Sandwich is an off-label technique that uses parallel stent grafts to treat aortoiliac aneurysms. The purpose of this experimental study is to evaluate the conformability and juxtaposition of stent grafts combinations used in this technique through in-vitro mechanical evaluation, computed tomography (CT) analyses, and a controlled pulsatile flow system.
Methods:
The combinations of two Viabahn® ("V-V") or Viabahn® and Excluder® iliac extension ("V-E") were analysed using CT imaging with measurement of the gutter area by two independent analysts before and after balloon angioplasty. In a second phase, the parallel stent combinations were also evaluated using CT imaging after being implanted in the aortic aneurysm model with a pulsatile flow system with controlled temperature, viscosity, and density.
Results:
The "V-E" group had a better conformability when compared to the "V-V" group, ensuring smaller gutter areas (0.0064 cm2 ± 0.01 vs. 0.0228 cm2 ± 0.03, p < 0.001). Post dilatation with two non-compliant balloons resulted in enlargement of the gutter area (Area A, p 0.065; Area B, p 0.071). Conversely, post dilatation with a non-compliant balloon for the internal iliac component and a compliant balloon for the external iliac device reduced the gutter area (Area A, p 0.008; Area B, p 0.010).
Conclusion:
The combination of Viabahn® and Excluder® iliac extension device ("V-E") had a smaller gutter area compared to two Viabahn® parallel stents for the Iliac Sandwich Technique. Post dilatation using a non-compliant balloon for the internal iliac device and a compliant balloon for the external iliac provided superior conformability and juxtaposition.
More Related Videos
06:09Femoral Vascular Graft Implantation in a Swine Model to Test Small-Diameter Vascular Grafts
Published on: July 8, 2025
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023