Use of Multiple Irradiations and Reference Materials as Comparators in Quality Control of Neutron Activation Analysis
A N Garg1,2, Vivek Singh3,4, R P Choudhury5,6
1Department of Chemistry, Indian Institute of Technology, Roorkee, 247667, India. amarnath943@yahoo.com.
Abstract:
We propose that several different reactor irradiation times followed by assaying of activity for differential counting periods may be employed for quality control (QC) of neutron activation analysis (NAA) data of biological samples. It is also recommended that three to four reference materials (RMs) of similar matrix but from different agencies such as National Institute of Standards and Technology (NIST, USA), International Atomic Energy Agency (IAEA, Vienna), Institute of Nuclear Chemistry and Technology (INCT, Poland), and National Institute of Environmental Studies (NIES, Japan) including a synthetic multielemental primary standard should always be analyzed simultaneously along with the samples. Finally, the mean ± σ values so obtained may be considered as more reliable after statistical analysis. Our analytical data for Na and K in three RMs (SRM 1572, CRM H-9, and Bowen's Kale) using different irradiation periods of 15 m, 2 h, 6 h, and 1 day were comparable with the certified values within error range of + 0.2 to - 2.7%. We report our data for 20 elements in two candidate RMs corn flour (INCT-CF-3) and soya bean flour (INCT-SBF-4) from the INCT, Poland, where Z-score values for most elements are in reasonable range of certified values.