Related Experiment Video
Updated: Sep 5, 2025

Author Spotlight: Integrating Alveolar-Capillary Reserve Measurements in Exercise Adaptation and Therapeutic Strategies
Published on: February 2, 2024
A Bench Comparison of the Effect of High-Flow Oxygen Devices on Work of Breathing
Claude Guérin1, Martin Cour2, Florian Degivry3
1Medicine Intensive-Réanimation Hôpital Edouard Herriot, Lyon, France; Université de Lyon, Lyon, France; and Institut Mondor de Recherches Biomédicales, INSERM 955, CNRS ERL 7000, Créteil, France. claude.guerin@chu-lyon.fr.
Background:
Oxygen therapy via high-flow nasal cannula (HFNC) has been extensively used during the COVID-19 pandemic. The number of devices has also increased. We conducted this study to answer the following questions: Do HFNC devices differ from the original device for work of breathing (WOB) and generated PEEP?
Methods:
Seven devices were tested on ASL 5000 lung model. Compliance was set to 40 mL/cm H2O and resistance to 10 cm H2O/L/s. The devices were connected to a manikin head via a nasal cannula with FIO set at 0.21. The measurements were performed at baseline (manikin head free of nasal cannula) and then with the cannula and the device attached with oxygen flow set at 20, 40, and 60 L/min. WOB and PEEP were assessed at 3 simulated inspiratory efforts (-5, -10, -15 cm H2O muscular pressure) and at 2 breathing frequencies (20 and 30 breaths/min). Data were expressed as median (first-third quartiles) and compared with nonparametric tests to the Optiflow device taken as reference.
Results:
Baseline WOB and PEEP were comparable between devices. Over all the conditions tested, WOB was 4.2 (1.0-9.4) J/min with the reference device, and the relative variations from it were 0, -3 (2-4), 1 (0-1), -2 (1-2), -1 (1-2), and -1 (1-2)% with Airvo 2, G5, HM80, T60, V500, and V60 Plus devices, respectively, (P < .05 Kruskal-Wallis test). PEEP was 0.9 (0.3-1.5) cm H2O with Optiflow, and the relative differences were -28 (22-33), -41 (38-46), -30 (26-36), -31 (28-34), -37 (32-42), and -24 (21-34)% with Airvo 2, G5, HM80, T60, V500, and V60 Plus devices, respectively, (P < .05 Kruskal-Wallis test).
Conclusions:
WOB was marginally higher and PEEP marginally lower with devices as compared to the reference device.
More Related Videos
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
06:11Employing the Forced Oscillation Technique for the Assessment of Respiratory Mechanics in Adults
Published on: February 9, 2022
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...