David Nash1, Shagun Juneja2, Antony L Palmer2
1Department of Medical Physics, Portsmouth Hospitals University NHS Trust, Portsmouth, UK; Division of Cancer Sciences, Faculty of Biology, Medicine and Health, University of Manchester, Manchester, UK.
Purpose:
Adaptive radiotherapy relies of rapid re-contouring, online more so than offline. Intra-patient contour propagation via non-rigid registration offers a solution but can be of limited accuracy. However, the dosimetric significance of the inaccuracies is unknown. Here we evaluate the dosimetric reliability of contours generated by different commercially-available software packages.
Method:
Planning CT contours for ten head and neck cancer patients were propagated via five commercial packages to five CBCT scans acquired throughout treatment. The treatment plan was recalculated on each of the CBCTs for each set of propagated contours, and DVH parameters extracted for the spinal cord, brainstem, parotids and larynx. The propagated contours were compared to two gold standard contours: contours manually outlined and a consensus STAPLE contours generated from the propagated contours. Geometrical similarity was evaluated using mean distance to agreement (mDTA), Hausdorff distance, centroid agreement and Dice similarity coefficient. Dosimetric reliability was assessed against clinical constraints and comparing via the intraclass correlation coefficient (ICC).
Results:
All propagated contours were similar to the STAPLE (mDTA < 1.0 mm) whilst larger differences were seen for the manual contours (mDTA < 3.0 mm). The dosimetric comparison showed that the propagated contours gave excellent dose estimates for most organs. The spinal cord reliability was moderate (ICC > 0.66).
Conclusions:
Large differences in geometric metrics rarely had a statistically significant impact on DVH parameters for the OARs studied. For that reason, propagated contours on treatment CBCT images are suitable for estimating dose to the OARs.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography