Integrating moving average control procedures into the risk-based quality control plan in small-volume medical
Vera Lukić1, Svetlana Ignjatović2,3
1Department of Laboratory Diagnostics, Railway Healthcare Institute, Belgrade, Serbia.
Introduction:
The modern approach to quality control (QC) in medical laboratories implies the development of a risk-based control plan. This paper aims to develop a risk-based QC plan for a laboratory with a small daily testing volume and to integrate the already optimized moving average (MA) control procedures into this plan.
Materials And Methods:
A multistage bracketed QC plan for ten clinical chemistry analytes was made using a Westgard QC frequency calculator. Previously, MA procedures were optimized by the bias detection simulation method.
Results:
Aspartate aminotransferase, HDL-cholesterol and potassium had patient-risk sigma metrics greater than 6, albumin and cholesterol greater than 5, creatinine, chlorides, calcium and total proteins between 4 and 5, and sodium less than 4. Based on the calculated run sizes and characteristics of optimized MA procedures, for 6 tests, it was possible to replace the monitoring QC procedure with an MA procedure. For the remaining 4 tests, it was necessary to keep the monitoring QC procedure and introduce MA control for added security.
Conclusion:
This study showed that even in a laboratory with a small volume of daily testing, it is possible to make a risk-based QC plan and integrate MA control procedures into that plan.
Related Concept Videos
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Introduction to Statistical Process Control
Quality Assurance
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
The R Chart
R charts are pivotal for pinpointing shifts in process variability. Stability is indicated when all data points remain within the defined upper and lower...