High frequency (1000 Hz) tympanometry in six-month-old infants
Sreedevi Aithal1, Joseph Kei2, Venkatesh Aithal3
1Department of Audiology, Townsville University Hospital, Queensland, Australia; Hearing Research Unit for Children, School of Health & Rehabilitation Sciences, University of Queensland, Australia; Department of Speech Pathology, School of Rehabilitation Sciences, James Cook University, Australia.
Objectives:
High frequency tympanometry (HFT) using a 1000 Hz probe tone is recommended for infants from birth to six months of age. However, there is limited normative HFT data outside the newborn period. The objective of this study was to describe HFT data in healthy six-month-old infants.
Methods:
HFT and distortion product otoacoustic emission (DPOAE) tests were performed on 168 six-month-old full-term healthy infants. Ears that passed DPOAEs and had a single-peaked tympanogram were included for analysis. The tympanometric measures included in the normative HFT data were tympanometric peak pressure (TPP), peak compensated static admittance (Ytm) and tympanometric width (TW).
Results:
A total of 118 ears from 118 infants who passed DPOAE and had single-peaked tympanograms were included in the analysis. Normative data were presented for TPP, Ytm and TW. A comparison of the present study with studies on neonates and younger infants revealed significantly higher mean Ytm and lower mean TPP for six-month-old-infants.
Conclusion:
Significant differences in HFT findings between neonates and six-month-old infants suggest a developmental trend and confirm the need for separate age-appropriate norms for the tympanometric measures. Normative HFT data described in the present study may provide useful information for optimizing the diagnosis of conductive conditions in six-month-old infants.