A Probing Device for Quantitatively Measuring the Mechanical Properties of Soft Tissues during Arthroscopy
Published on: May 1, 2020
5.6K
Retropubic trocar modified with a load cell to verify contact with pubic bone.
Md A Arif1, Antonis P Stylianou1, Austin Bachar2
1School of Computing and Engineering, University of Missouri Kansas City, MO.
Surgery
|July 12, 2022
Summary
A modified trocar with a load cell helps surgeons avoid vital injuries during midurethral sling surgery by confirming bone contact. This tool demonstrated reliability and accuracy in simulations, aiding novice surgeons and improving patient safety.
Area of Science:
- Urology
- Surgical Simulation
- Medical Device Innovation
Background:
- Midurethral sling surgery carries risks of vital injuries due to the blind nature of trocar passage.
- Monitoring trocar-bone contact is crucial but challenging for teaching surgeons.
- A novel retropubic trocar equipped with a load cell was developed to detect bone contact during surgery.
Purpose of the Study:
- To evaluate the biofidelity and test-retest reliability of a modified retropubic trocar on a 3D surgical simulator.
- To assess the trocar's ability to differentiate between on-bone and off-bone movements.
- To determine the potential of this device in enhancing surgical training and patient safety.
Main Methods:
- The modified trocar was tested on a 3D midurethral sling surgery simulator using expert and novice surgeons.
- Surgeons performed retropubic trocar passage on a physical pelvic floor model.
- Performance metrics including force variables were analyzed and compared between experts and novices, and between cadaveric and model testing.
Main Results:
- The modified trocar demonstrated adequate biofidelity, with similar force readings between cadaveric and model testing (P = .62).
- High test-retest reliability was observed, with similar force variables across repeated trials (P = .30).
- Expert surgeons showed significantly different force application patterns compared to novices, indicating the simulator's ability to discern skill levels.
Conclusions:
- The modified trocar exhibits high test-retest reliability and adequate biofidelity for surgical simulation.
- This innovative device can assist novice surgeons in maintaining correct trocar placement during midurethral sling procedures.
- The trocar has potential applications in both surgical simulation and the operating room to enhance surgical safety and monitoring.