Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Harmonization based on quantitative analysis of standardized uptake value variations across PET/CT scanners: a
Abbas Monsef1,2, Mohammad Reza Ay1,2, Peyman Sheikhzadeh1,3
1Department of Medical Physics and Biomedical Engineering.
Harmonization techniques significantly reduced variations in standardized uptake value (SUV) measurements across different PET/CT scanners. This improves quantitative accuracy and reproducibility in nuclear medicine imaging.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Quantitative Analysis
Background:
- Standardized uptake value (SUV) quantification in PET/CT can vary significantly between scanners.
- Inter-system variability hinders accurate comparison of imaging results.
Purpose of the Study:
- To measure and reduce variations in SUV measurements across different PET/CT scanners.
- To harmonize quantitative PET/CT imaging for improved reproducibility.
Main Methods:
- Utilized the NEMA IEC phantom across three PET/CT scanners.
- Applied Gaussian filters with tuned FWHM for SUV harmonization.
- Assessed inter-site SUV variations using digital reference objects and EANM/EARL recommendations.
- Performed test-retest analysis for repeatability.
Main Results:
- Harmonization significantly decreased inter-site SUV differences for SUVmean, SUVmax, and SUVpeak (e.g., SUVmean from 17.3% to 4.8%).
- P values <0.05 confirmed statistical significance of harmonization.
- Image quality was preserved, with contrast-to-noise ratio >2.8% in the smallest phantom lesion.
Conclusions:
- Harmonization effectively reduces SUV measurement variability across PET/CT scanners.
- The applied method improves the reproducibility of quantitative PET/CT imaging.
- Image quality is maintained post-harmonization, ensuring reliable diagnostic information.