Abstract:
The determination of ammonium ions (NH4+) is of significance to environmental, agriculture, and human health. Potentiometric NH4+ sensors based on solid-contact ion selective electrodes (SC-ISEs) feature point-of-care testing and miniaturization. However, the state-of-the-art SC-ISEs of NH4+ during the past 20 years strongly rely on the organic ammonium ionophore-based ion selective membrane (ISM), typically by nonactin for the NH4+ recognition. Herein, we report a Prussian blue analogue of copper(II)-hexacyanoferrate (CuHCF) for an ISM-free potentiometric NH4+ sensor without using the ionophores. CuHCF works as a bifunctional transducer that could realize the ion-to-electron transduction and NH4+ recognition. CuHCF exhibits competitive analytical performances regarding traditional nonactin-based SC-ISEs of NH4+, particularly for the selectivity toward K+. The cost and preparation process have been remarkably reduced. The theoretical calculation combined with electrochemical tests further demonstrate that relatively easier intercalation of NH4+ into the lattices of CuHCF determines its selectivity. This work provides a concept of the ISM-free potentiometric NH4+ sensor beyond the nonactin ionophore through a CuHCF bifunctional transducer.
Related Concept Videos
Potentiometry: Membrane Electrodes
Amperometry: Overview
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Overview
Potentiometric Titration: Overview
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential...