Related Experiment Video
Updated: Sep 4, 2025

ELIME Enzyme Linked Immuno Magnetic Electrochemical Method for Mycotoxin Detection
Published on: October 23, 2009
Food allergen control: Tropomyosin analysis through electrochemical immunosensing
Ricarda Torre1, Maria Freitas1, Estefanía Costa-Rama2
1REQUIMTE/LAQV, Instituto Superior de Engenharia do Porto, Instituto Politécnico do Porto, Rua Dr. Ant ónio Bernardino de Almeida 431, 4200 -072 Porto, Portugal.
Abstract:
Regulations of the EU obliges the indication of the presence of allergens on food labels. This work reports the development of an electrochemical immunosensor to determine tropomyosin (TPM) - a major shellfish allergen - prevailing in the muscles of crustacean species. Two linear ranges between the signal and TPM concentration were obtained: between 2.5 and 20 ng mL-1 and between 30 and 200 ng mL-1, with a lowest limit of detection of 0.47 ng mL-1. The selectivity of the optimized immunoassay, tested with other food allergens (e.g., Cyp c 1, a fish allergen), assures the effective detection of TPM, enabling successful control of foodstuff labelling. Several (12) foods, containing high and low TPM concentrations and TPM-free samples, were analysed using the sensor. A conventional ELISA kit and recovery assays were used to evaluate the accuracy of the results.