AN INTER-COMPARISON OF PASSIVE DOSEMETERS FOR TERRESTRIAL GAMMA RADIATION MEASUREMENTS
Yanqiu Ding1, Kaiyi Wang1, Pin Gao1
1Key Laboratory of Radiological Protection and Nuclear Emergency, China CDC, National Institute for Radiological Protection, Chinese Center for Disease Control and Prevention, Beijing 100088, China.
Abstract:
To evaluate the measurement method of passive dosemeters for terrestrial gamma radiation measurements. An inter-comparison of passive dosemeters for terrestrial gamma radiation measurements was organized alternately by the Japan Chemical Analysis Center, National Institute for Radiological Protection of Chinese Center for Disease Control and Prevention and Radiation Monitoring Technical Center of Ministry of Environmental Protection every 2 years. The procedures and exposure conditions of the field-exposure and reference irradiation test were provided by the organizer. During 2019-2020, the En values of Field-1 and Field-2 in three laboratories were lower than 1.0 and assigned 'acceptable' according to inter-comparison criteria. After the correction of the fading effect of dosemeter material, the En values of reference irradiation test (0.301 mGy, 1.000 mGy) in three laboratories were lower than 1.0 and assigned 'acceptable'.