Assessment of image quality in photon-counting detector computed tomography of the wrist - An ex vivo study
Nina Kämmerling1, Mårten Sandstedt2, Simon Farnebo3
1Department of Radiology, and Department of Health, Medicine and Caring Sciences, Linköping University, Linköping, Sweden.
Purpose:
The aim of this study was to evaluate the effect of reconstruction parameters on image quality in wrist imaging using photon-counting detector CT (PCD-CT) and to compare the results with images from an energy-integrating detector CT (EID-CT).
Methods:
Twelve cadaveric wrist specimens were examined using a prototype PCD-CT and a clinical EID-CT using similar radiation dose. Reconstruction parameters were matched between scanners. Also, sharper reconstruction kernels, a larger matrix size, and smaller slice thicknesses were evaluated for PCD-CT. Image noise, contrast-to-noise ratio (CNR) and image sharpness in trabecular structures were quantitatively measured. Image quality with respect to the visibility of cortical and trabecular bone structures was assessed by six radiologists using visual grading methods.
Results:
Images obtained with PCD-CT had lower noise (42.6 ± 3.9 HU vs 75.1 ± 6.3 HU), higher CNR (38.9 ± 4.5 vs 19.0 ± 2.4) and higher trabecular sharpness (63.5 ± 6.0 vs 53.7 ± 8.5) than those obtained with EID-CT using similar scan and reconstruction parameters (p < 0.001). The image sharpness in trabecular structures was further improved by using sharper kernels, despite higher noise levels. Radiologists had a strong preference for PCD-CT images both in terms of spatial resolution and suitability for bone imaging. Visual grading analysis showed an improved visibility of cortical bone, trabeculae and nutritive canals (p < 0.005).
Conclusion:
PCD-CT offers improved image quality regarding bone structures in the wrist relative to EID-CT systems, particularly when sharper reconstruction kernels, smaller slice thickness and a larger image matrix size are used.
More Related Videos
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET