The impact of ASiR-V on abdominal CT radiation dose and image quality - A phantom study
Svea Deppe Mørup1, John Stowe2, Martin Weber Kusk3
1Health Sciences Research Centre, UCL University College, Campus Odense: UCL Erhvervsakademi og Professionshojskole, Campus Niels Bohrs Alle, Radiography, Odense, Denmark; Cardiology Research Department, Odense University Hospital, Svendborg, Denmark; Radiography and Diagnostic Imaging, School of Medicine, University College Dublin, Ireland.
Introduction:
To investigate how ASiR-V and kVp changes in Computed tomography (CT) affect radiation dose and image quality, when using automatic tube current modulation (ATCM) for different sized phantoms.
Methods:
A liver-phantom with two different liver inserts (QRM, Moehrendorf, Germany), with extension rings, representing fat, were additionally applied to the phantom to simulate patients of different sizes (small: 30cm diameter, medium: 35cm and large: 40cm). Abdominal scans were performed on a 256 slice CT scanner (GE Healthcare, Milwaukee, WI, USA), with consistent pitch (0.992), rotation time (0.5s), slice thickness (0.625mm) and collimation (80mm), while other parameters were varied (kVp: 80/100/120/140; Noise Index: 13/22; mA interval 80-720, ASiR-V: 30/60/100%). CTDI and DLP was recorded for each scan and image quality was assessed using objective metrics in predefined anatomic areas (HU and noise). Radiation dose and image quality metrics were compared between protocols.
Results:
CTDI decreased by 80% from ASIR-V 30% to ASiR-V 100% for prescribed NI 13, and by 79% for the prescribed NI of 22. For 100% ASiR-V and a prescribed NI of 22 the CTDI remained the same regardless of phantom size for the different kVp settings. Pairwise comparison revealed significant differences in CTDI (p < 0.0001) for all combinations of prescribed NI and ASIR-V levels, except the difference between ASIR-V levels of 30 and 60%, with a prescribed NI of 13 (p = 0.124). When data from the three phantom sizes were combined, increasing ASIR-V from 30-100%, resulted in noise decreases of 22% for NI of 13 and by 8% for NI of 22. Notably, image quality in the low contrast area of the liver insert was impaired when the large phantom was scanned with 100% ASiR-V and either 80/100kVp (NI 22), because of the large reduction in tube current applied (down to 80 mA).
Conclusion:
Substantial radiation dose reductions (up to 80%) resulted from increasing ASiR-V levels. However, image quality deteriorates when 100% ASiR-V is applied due to low applied tube current by the ATCM.
More Related Videos
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
11:05Tissue-simulating Phantoms for Assessing Potential Near-infrared Fluorescence Imaging Applications in Breast Cancer Surgery
Published on: September 19, 2014
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
X-ray Imaging