Updated: Sep 4, 2025

Electrochemiluminescence Assays for Human Islet Autoantibodies
Published on: March 23, 2018
A multicenter performance evaluation of the new Elecsys Vitamin D total III assay versus reference isotope dilution
Peter Findeisen1, Marta Leis2, Garnet Bendig3
1MVZ Labor Dr. Limbach & Kollegen, Heidelberg, Germany.
Background:
Vitamin D deficiency/insufficiency and toxicity are worldwide issues; thus, accurate diagnostic assays are required to measure vitamin D. We evaluated the performance of the new Elecsys® Vitamin D total III assay (Roche Diagnostics International Ltd).
Methods:
Repeatability and intermediate precision of the Elecsys Vitamin D total III assay (cobas e 601 analyzer) were evaluated at three sites using five human serum pools (HSPs) and two PreciControls (five-day model, one reagent lot [CLSI-EP05-A3]) and compared against prespecified acceptance criteria. A serum verification panel, with reference isotope dilution liquid chromatography tandem mass spectrometry (ID-LC-MS/MS) values, was used for comparator assay/concordance studies at two sites, assessed using unweighted Deming regression. Testing of serum vs. plasma on the Elecsys assay was conducted at one site using samples from healthy adults; assessed using Passing-Bablok regression.
Results:
Repeatability (HSP1 [16.8-18.4 ng/ml], SD 0.87-1.07; HSP5 [94.5-98.0 ng/ml], CV 1.58%-2.76%) and intermediate precision (HSP1, SD 1.14-1.77; HSP5, CV 2.00%-4.13%) met acceptance criteria across sites. Agreement was observed between the Elecsys assay and (i) the ID-LC-MS/MS verification panel (slope, 0.936-1.01; Pearson's r, 0.960-0.986) and (ii) comparator assays (slope, 0.921-1.15; Pearson's r, 0.958-0.982). The Elecsys assay correctly assigned the highest combined percentage of samples to deficient (100%) and insufficient (89.5%) vitamin D categories vs. comparator assays and demonstrated comparable performance in serum and plasma (y = 0.103 + 0.984x).
Conclusions:
The Elecsys Vitamin D total III assay demonstrated good analytical performance and compared favorably with other assays, supporting its use in clinical practice.
More Related Videos
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...