The impact of vertical off-centering on image noise and breast dose in chest CT with organ-based tube current
Teemu Mäkelä1, Mika Kortesniemi2, Touko Kaasalainen2
1HUS Diagnostic Center, Radiology, University of Helsinki and Helsinki University Hospital, Helsinki, Finland; Department of Physics, University of Helsinki, Finland.
Purpose:
To determine the effects of patient vertical off-centering when using organ-based tube current modulation (OBTCM) in chest computed tomography (CT) with focus on breast dose.
Materials And Methods:
An anthropomorphic adult female phantom with two different breast attachment sizes was scanned on GE Revolution EVO and Siemens Definition Edge CT systems using clinical chest CT protocols and anterior-to-posterior scouts. Scans with and without OBTCM were performed at different table heights (GE: centered, ±6 cm, and ± 3 cm; Siemens: centered, -6 cm, and ± 3 cm). The dose effects were studied with metal-oxidesemiconductor field-effect transistor dosimeters with complementary Monte Carlo simulations to determine full dose maps. Changes in image noise were studied using standard deviations of subtraction images from repeated acquisitions without dosimeters.
Results:
Patient off-centering affected both the behavior of the normal tube current modulation as well as the extent of the OBTCM. Generally, both OBTCM techniques provided a substantial decrease in the breast doses (up to 30% local decrease). Lateral breast regions may, however, in some cases receive higher doses when OBTCM is enabled. This effect becomes more prominent when the patient is centered too low in the CT gantry. Changes in noise roughly followed the expected inverse of the change in dose.
Conclusions:
Patient off-centering was shown to affect the outcome of OBTCM in chest CT examination, and on some occasions, resulting in higher exposure. The use of modern dose optimization tools such as OBTCM emphasizes the importance of proper centering when preparing patients to CT scans.
More Related Videos
08:34Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...